Optical method for the calibration of raster scales
| Author | Affiliation |
|---|---|
Grattan, Kenneth Thomas Victor | City University, London |
| Year | Volume | Issue | Start Page | End Page |
|---|---|---|---|---|
2002 | 32 | 1 | 23 | 29 |
| URI | Access Rights |
|---|---|
| https://hdl.handle.net/20.500.14911/130814 | |
| Straipsnis | Viso teksto dokumentas (atviroji prieiga) / Full Text Document (Open Access) |
High quality techniques for the calibration of linear and circular raster scales used in photoelectric transducers and the information-measuring systems of a range of machines and scientific instruments is of particular importance, enabling them to give a more accurate performance. The most significant methods of measurement in this area are discussed in this paper, and a new approach to their calibration is presented. Most methods used for optical measurement scales apply a standard measure of angle or length and use photoelectric devices for the readout. A new half circle method for the calibration of circular scale is described. A novel idea for the measurement of linear and circular raster scales is described, given by the determination of a bias from a straight line and from a circle, respectively. This idea is implemented by applying a Moiré fringe conjunction technique and the modification of Moiré fringe parameters resulting, expressed by discrete functions, is discussed.