Skaitmeninių nivelyrų NA3003 kolimacijos paklaidos tyrimai
| Year | Volume | Issue | Start Page | End Page |
|---|---|---|---|---|
2001 | XXVII | 1 | 36 | 39 |
One of the most important digital levels errors, which influences the accuracy, is the collimation error. Digital levels NA2000/NA3000 have collimation errors, which include the absolute error absColl and the variable error CollDif. During measurements digital levels by electronical reading from the code rods correct automatically the readings due to collimation error absColl, if it was defined and put into the memory of the instrument. If absColl is bigger than 20", it is suggested to regulate the position of the threads grid. The investigation and calculation of the CollDif error is presented in the paper. On August 1, 2000, the analysis of the collimation error change of the two levels NA3003 was done by using permanently equipped code rods and tribrachs. It was determined that the change of the CollDif seeks 4-8", if the temperature change is 5ºC. So the differences of the distances to the back and forward rods should not exceed 0,5 m. CollDif of digital levels have the systematic influence on levelling results, because the firmware of these levels allows the levelling according to the program BFFB.[...].